Patient Isolation Carrier

INTRODUCTION

WS-N-01 is a compact patient isolation carrier type transport equipment, designed to prevent secondary infection from confirmed or suspected cases of infectious diseases during transportation.

CHARACTERISTICS

 Category
  Specifications
Model 
WS-N-01  
Airflow rate (m3/h)     
35
Dimensions (mm) 
1930L x 550W x 440H
Power 
DC 12V 
Pressure (Pa)  
up to -15 
Noise level (dB) 
58
HEPA filter  
99.999%, 0.3 µm 
Weight (kg)
10 

CERTIFICATION 

CE

In accordance to Regulation (EU) 2017/745, WS-N-01 Medical device class 1 
has been designed to comply with the requirement of the following Harmonized Standard:
EN ISO 13485:2016
EN 1789:2007 + A1:2010
EN 61000-6-3:2007/A1:2011/AC:2012 

KGMP 

Certified, that Wosem Co., Ltd. manufacturer complies with Korea Good Manufacturing Practices of Medical Devices.

SPECIFICATIONS

Airframe Design

● CT Scannable (It fits and doesn’t interfere with CT machines)
● Low cost
● Light weight
● Simple installation (Less than 5 minutes)
● Easy to disassemble and stock (carry-on bag enclosed)

Features

● Check and control pressure level and battery remains through the display screen
● H14 class HEPA filter attached to keep the exhaust air safe
● Easy to replace used HEPA filter
● Waterproof zippers
● Built-in gloves with sleeves
● Battery remains for 6 hours and can be charged while in use
● Additional 2 hours usage with auxiliary 8 AA size batteries

Patient Isolation Wheelchair

INTRODUCTION

NPW-N-01 The Patient Isolation Wheelchair is a reliable type of transport equipment, designed to prevent secondary infection from confirmed or suspected cases of infectious diseases during transportation on paved roads. Its solid frame and user-friendly design, including large transparent windows and display screen have received good reviews from hospital personnel in Korea and abroad.

CHARACTERISTICS

 

 Category
Specifications 
Model 
NPW-N-01 
Air Change rate (m3/h) 
35 
Dimensions (mm) 
620W x 1350D x 1660H
Chamber dimensions (mm) 
580W x 1000D x 1445H
Power 
DC 12V 
Pressure (Pa)
from -2,5 to -15
Noise level (dB)
65
HEPA filter
99.999%, 0.3 µm

CERTIFICATIONS

CE

In accordance to Regulation (EU) 2017/745, WS-N-01 Medical device class 1 
has been designed to comply with the requirement of the following Harmonized Standard:
EN ISO 13485:2016
EN 1789:2007 + A1:2010
EN 61000-6-3:2007/A1:2011/AC:2012 

KGMP 

Certified, that Wosem Co., Ltd. manufacturer complies with Korea Good Manufacturing Practices of Medical Devices.

SPECIFICATIONS

Observation windows

● Large transparent window enables the medical staffs to check patient’s condition during transportation.
● First Aid Zippers attached on the side windows for IV injection if needed

Features

● Solid frame : STS304(25mm) 
● Check and control pressure level and battery remains through the display screen
● H14 class HEPA filter attached to keep the exhaust air safe
● Easy to replace used HEPA filter
● Waterproof zippers
● Battery remains for 6 hours and can be charged while in use
● Additional 2 hours usage with auxiliary 8 AA size batteries

Energy Recovery Ventilation & Emergency Mode

INTRODUCTION

WS-NHE-01 This device function as Energy Recovery Ventilation (ERV) in ventilation mode. In isolation mode it prevents secondary infection by forming negative pressure in a designated isolation room when self-isolation is required, by cleaning air and preventing viruses from spreading.

GENERAL FEATURES

● PRE filter – grabs large particle dusts
● HEPA filter – grabs fine dust and high-risk pathogens
● Control Pad – installed outside the room to set or check operation mode and negative pressure status inside the room.
● Precise control
● Automatic operation by CO2 Sensor to maintain fresh air quality.
● Air supply mode – supplies fresh air into the house.
● Air exhaust mode – exhausts the inside air out of the house.

SPECIFICATION

Model
WS-NHE-01
WS-NHE-02
Dimensions (mm)
700W*700D*250H     
750W*800D*350H     
Power Consumption
120W
360W
Heat recovery efficiency  
<70%
<70%
Supply air volume
150 m³/h
500 m³/h
Exhaust air volume
150 m³/h
500 m³/h
PRE Filter
5T 300*300*21H
5T 310*350*21H
HEPA Filter
15T 300*300*21H
15T 310*350*21H
Motor
BLDC
Controller
Wall-mounted / Network System

Ceiling Negative Pressure Unit

INTRODUCTION

Ceiling Negative Pressure Unit can be installed on the ceiling of the patient rooms to convert it into a temporary Negative Pressure Isolation Room(NPIR) in case of an emergency case related to infectious diseases. After the use of NPIR is done, our ceiling negative pressure units can be used for ventilation and air purification(optional).It takes much less time and cost then constructing a typical NPIR.

CHARACTERISTICS

COVER RANGE

Ceiling Negative Pressure Unit applicable for up to 100㎥ ±15% room

GENERAL FEATURES

Aerodynamic design presents significantly less noise and vibration compared to other similar products PRE filter grabs large particle dusts HEPA filter grabs fine dust and high-risk pathogens Digital display installed outside the room to check the negative pressure status inside the room. Air volume adjustable by BLDC motor (0 ~ 25 CMM)

NETWORK SYSTEM (Optional)

Built-in RS485 network system allows you to check and control the negative pressure status and filter replacement cycle of the rooms at the central management room.
 

Portable Negative Pressure Unit

INTRODUCTION

Portable Negative Pressure Unit can be easily installed in a single or multiple patient room to convert it into a temporary. Negative Pressure Isolation Room(NPIR) in case of an emergency case related to infectious diseases. After the use of NPIR is done, our portable negative pressure units can be easily removed and stocked in storage. It takes much less time and cost then constructing a typical NPIR. 

CHARACTERISTICS

COVER RANGE

WS-M-500 applicable for up to 30㎥ ±15% room
WS-M-2400-2 applicable for up to 100㎥ ±15% room

GENERAL FEATURES

Aerodynamic design presents significantly less noise and vibration compared to other similar products
PRE filter grabs large particle dusts
HEPA filter grabs fine dust and high-risk pathogens
Digital display installed outside the room to check the negative pressure status inside the room.
Air volume adjustable by BLDC motor (0 ~ 25 CMM)

NETWORK SYSTEM (Optional)

Built-in RS485 network system allows you to check and control the negative pressure status and filter replacement cycle of the rooms at the central management room